Agilent E3633A Repair (In Progress)

Created: November 5, 2026

Background

I ended up purchasing a non-working Agilent E3633A DC power supply off of eBay as I suspected I would in my previous Agilent E3631A repair post. The Agilent E3633A is a single-output programmable DC power supply with two range options (0-8V/20A or 0-20V/10A), meaning this is a 200W power supply. The unit itself is in really good cosmetic shape (see image from listing below), but the seller’s description said this:

For parts or repair. The output does not respond when the voltage is increased, even with OVP and OCP turned off. Sold AS-IS for parts.

I have been enjoying the process of troubleshooting Agilent equipment lately so I thought I would take on the challenge. I submitted an offer to the seller and it was accepted. Ended up costing me around $180 with shipping. Not horrible for a 200W power supply!

Listing from eBay

Arrival

The unit was packaged extremely well, which is unusual for me lately! When I got the unit, I powered it on to see if I could replicate what the seller described. I turned the output on and immediately noticed that the power supply was stuck in constant current (CC) mode. I measured the output current through my DMM and it was showing about half of the value displayed on the screen. My first thought was that the unit must have a short-circuit somewhere. I’m not going to lie, I was a little worried since there were not many repair posts online for this unit.

Disassembly

To start taking the unit apart I removed the four Torx screws on the back of the unit and the four Phillips on the sides of the unit (two on each side). From there the back panel should be removed and the shell will slide backwards. To help with troubleshooting I also removed the front panel from the main unit by removing the four Phillips screws underneath the rubber bumper (two on each side). I noticed these had Blue Loctite. Now the front panel can be put aside so there is more room to inspect the unit. Next, the transformer can come out by removing the four Phillips screws (with more Blue Loctite), however it would be wise to first disconnect the wire leading to P11. Now the transformer can be lifted and placed behind the unit on its side without disconnecting any more wires. I used jumper wires to connect the transformer back to P11. To take it a step further the whole board can be removed from the chassis, but I did not do so.

Disassembled Unit

Troubleshooting

First, I put the unit on its back with books used to prop the back up. There are quite a few different test points on the back with silkscreen pointing out what they’re for. I went around and measured each point only to find them close to their nominal values. After all of those checked out I decided to look at the schematic in more detail. I do not have much experience with analog design so I like to try to figure out what different parts of the circuit are doing. However, I sometimes struggle with this. Anyway, I decided to start at the banana jacks and work my way backwards. When tracing the output backwards I noticed a sense resistor. I had the power supply outputting 1A in CC mode. I measured the voltage across the sense resistor and it was 10mV. This checks out since the sense resistor is 10mOhm. I then decided to look between the sense resistor and the output jacks. There are a couple of diodes (CR41 and CR42) between the +/- and sense nodes. I turned the unit off and put my DMM in diode mode. To my surprise CR42 measured okay, but CR41 was shorted!

From here I removed CR41 with my desoldering gun, powered the unit back on, and the short was gone! The unit would now operate in constant voltage (CV) mode as well as constant current. I tested all the voltage and current settings. In the meantime I went ahead and ordered a few replacement parts. The service guide said they are RURG3040 diodes, but RURG3060s were installed in my unit. DigiKey has RURG3060s in stock but it seems that the RURG3040s are discontinued.

Final Thoughts

Once the new parts arrive I will solder them in place and test the unit again. I will update accordingly.